    Multifunction Calibrator MC30-3A


    Main features

    1. 5 1 / 2 LED digit displays output, which displays simultaneously by two ways, one bases on actual

    value, the other bases on percentage.

    2. Use the keyboard to control the increase or decrease of the output. Keyboard control volume is

    divided into 100% / N, 10% / N, 1% / N, 0.1% / N (N for 4, 5, 6, 10, 15).

    3. The regulator with external control can control the increase or decrease of output, there are

    two types of configuration which include keyboard and potentiometer, and they can converse

    with each other casually.

    4. Output range of AC, DC voltage is 0 ~ 1050V

    5. Output range of AC, DC current is 0 ~ 20A

    6. AC offers four kinds of output frequency: 50Hz, 60Hz (59.7Hz), 400Hz, 1000Hz, and uses crystal

    oscillator to ensure the accuracy and stability of frequency.

    7. Output overload is able to protect automatically, and reset manually.

    8. In order to reduce beat frequency influence of measured instrument, AC 50Hz output can also

    choose to synchronize their electricity.

    9. clamp-meter measurement: equipped with standard coil of our factory, 0 ~ 1000A current is

    measurable, and margin of error is ± 0.3%.


    Technical Parameters


    1. Stability: AC < 0.02% / 5 minutes of full range

    DC < 0.01% / 5 minutes of full range

    Instruments should be preheated for two hours for precision measurement

    2. Distortion of AC: <0.5%

    3. Ripple factor of DC: <0.05%( rated output)

    4. Accuracy of the output frequency: 50Hz, 60Hz (59.7Hz), 400Hz, 1KHz <0.1%

    (commercial power synchronization)

    5. The range and accuracy of the output voltage, current, as well as the medium resistance (23 ℃ ± 2 ℃,

    the value of output is more than 10% of the range)

    Output itemsOutput rangeRated outputMaximum outputAccuracy

    AC-DC voltage

    0~300mV20mA100mA

    basic error :DC:±(0.02% reading +0.03% Range )

    AC:±(0.02%reading

    +0.03% Range )

    1KHzAdditional error

    ±0.05%reading

    Additional error of voltage:

    ±0.02mV


    ACI:≤1mA not evaluate accuracy

    0~1V~3V50mA100mA
    0~5V~10V~30V100mA200mA
    0~50V~100V~300V50mA100mA
    0~250V~500V~1000V

    DC:20mA

    AC:40mA

    DC:40mA

    AC:60mA


    AC-DC current


    0~100mA

    DC:3V


    AC:36V

    0~1A

    DC:3V

    AC:12V

    0~20A1.5V
    medium resistance10Ω~24MΩ0.25W±0.2%+20mΩ
    P.S.(1)Working environment temperature exceeds 23 ℃ ± 2 ℃, every 10 ℃ changes, the additional error is less than basic errors of the range (2) 50A 1KHz error is ± (0.15% reading +0.05% range)

    Attach list

    6. Power supply features: AC power voltage 220V ± 10%, the frequency 50Hz ± 1Hz;

    7. Working environment: the temperature of working environment is 5 ℃ ~ 35 ℃, relative humidity <80%

    8. Working time: continuity

    9. Size: 180 × 480 × 480mm3

    10. Weight: about 24kg,

    Quality AC DC Multifunction Tester Calibration , 20A Multifunction Calibrator for sale